ADNI MRI CoreMRI Core News | |||||||||||||||||||
| Added: | |||||||||||||||||||
| > > |
Preprocessed Image Recalls - 06/05/08
| ||||||||||||||||||
Preprocessed Image Recalls - 10/3/07Please discard any preprocessed Scaled data from the subjects and scan dates below that were downloaded from LONI before 10/03/2007. | |||||||||||||||||||
| <<O>> Difference Topic ADNIMRICore (r1.7 - 04 Oct 2007 - KarenCrawford) |
ADNI MRI Core | ||||||||
| Changed: | ||||||||
| < < |
MRI Core News | |||||||
| > > |
MRI Core News | |||||||
| Changed: | ||||||||
| < < |
Duplicate Scaled Images | |||||||
| > > |
Preprocessed Image Recalls - 10/3/07Please discard any preprocessed Scaled data from the subjects and scan dates below that were downloaded from LONI before 10/03/2007.
Duplicate Scaled Images | |||||||
| A number of scaled images were uploaded in duplicate. They have now been deleted. All users should refrain from analyzing the following Image IDs if they were downloaded prior to 21 Sept 07: 45104, 38794, 45203, 46409, 47058, 34420, 48761, 60501, 60490, 42525 | ||||||||
| Added: | ||||||||
| > > |
| |||||||
| Changed: | ||||||||
| < < |
Philips Scanners Image Scaling | |||||||
| > > |
Philips Scanners Image Scaling | |||||||
| The ADNI MRI core has uncovered a potential concern about the accuracy of image scaling in the frequency encode direction on Philips 1.5T systems. This problem is currently under investigation. For the time being, we suggest that investigators use this particular data judiciously. This message will be updated as more information becomes available. | ||||||||
| Added: | ||||||||
| > > |
| |||||||
| Changed: | ||||||||
| < < |
ADNI MR Recall - NEW | |||||||
| > > |
ADNI MR Recall | |||||||
| Occasionally, the MR Core identifies preprocessed data through the Quality Control Process that is incorrect. In such cases, the MR Core, through LONI, will make an attempt to contact | ||||||||
| Line: 42 to 70 | ||||||||
|---|---|---|---|---|---|---|---|---|
| The MR Core found discrepancies in the ADNI phantom data analysis indicating the wrong gradient coefficient had been used to GW process these studies. The correct gradient was verified with the MR sites, the data was reprocessed and re uploaded to LONI. | ||||||||
| Added: | ||||||||
| > > |
| |||||||
| Masks | ||||||||
| Line: 50 to 79 | ||||||||
| ||||||||
| Changed: | ||||||||
| < < |
-- KarenCrawford - 21 Sept 2007 | |||||||
| > > |
-- KarenCrawford - 4 Oct 2007 | |||||||
| <<O>> Difference Topic ADNIMRICore (r1.6 - 21 Sep 2007 - KarenCrawford) |
ADNI MRI CoreMRI Core News | ||||||||
| Added: | ||||||||
| > > |
Duplicate Scaled Images A number of scaled images were uploaded in duplicate. They have now been deleted. All users should refrain from analyzing the following Image IDs if they were downloaded prior to 21 Sept 07: 45104, 38794, 45203, 46409, 47058, 34420, 48761, 60501, 60490, 42525 | |||||||
| Philips Scanners Image Scaling The ADNI MRI core has uncovered a potential concern about the accuracy of image scaling in the frequency encode direction on Philips 1.5T systems. This problem is currently under investigation. For the time being, we suggest that investigators use this particular data judiciously. This message will be updated as more information becomes available. | ||||||||
| Line: 46 to 50 | ||||||||
|---|---|---|---|---|---|---|---|---|
| ||||||||
| Changed: | ||||||||
| < < |
-- KarenCrawford - 07 May 2007 | |||||||
| > > |
-- KarenCrawford - 21 Sept 2007 | |||||||
| <<O>> Difference Topic ADNIMRICore (r1.5 - 15 Jun 2007 - KarenCrawford) |
ADNI MRI CoreMRI Core News | ||||||||
| Line: 35 to 35 | ||||||||
|---|---|---|---|---|---|---|---|---|
| ||||||||
| Deleted: | ||||||||
| < < |
||||||||
| The MR Core found discrepancies in the ADNI phantom data analysis indicating the wrong gradient coefficient had been used to GW process these studies. The correct gradient was verified with the MR sites, the data was reprocessed and re uploaded to LONI. | ||||||||
| Added: | ||||||||
| > > |
Masks | |||||||
| Added: | ||||||||
| > > |
Masks created by the MR Core as part of preprocessing are useful for performing N3 correction and are not brain masks. Therefore, the description of these masks has been updated to read "Intracranial Space" rather than "Brain." | |||||||
Topic permissions
| ||||||||
| <<O>> Difference Topic ADNIMRICore (r1.4 - 12 Jun 2007 - KarenCrawford) |
ADNI MRI CoreMRI Core News | ||||||||
| Added: | ||||||||
| > > |
Philips Scanners Image Scaling | |||||||
| The ADNI MRI core has uncovered a potential concern about the accuracy of image scaling in the frequency encode direction on Philips 1.5T systems. This problem is currently under investigation. For the time being, we suggest that investigators use this particular data judiciously. This message will be updated as more information becomes available. | ||||||||
| Added: | ||||||||
| > > |
ADNI MR Recall - NEW
Occasionally, the MR Core identifies preprocessed data through the Quality Control Process
that is incorrect. In such cases, the MR Core, through LONI, will make an attempt to contact
institutions that have downloaded data identified as incorrect. All investigators should
regularly check this site for recall notices and remove data previously downloaded for
analysis until corrected data has been posted.
For more information on MR recall advisories contact: Bret Borowski
borowski.bret@mayo.edu
Incorrect GE Gradients used for Gradient Warping Correction Please discard any preprocessed GW data from the subjects and scan dates below that were downloaded from LONI before 06/01/2007. Subject - Scan Date
| |||||||
| Added: | ||||||||
| > > |
The MR Core found discrepancies in the ADNI phantom data analysis indicating the wrong gradient coefficient had been used to GW process these studies. The correct gradient was verified with the MR sites, the data was reprocessed and re uploaded to LONI. | |||||||
Topic permissions | ||||||||
| <<O>> Difference Topic ADNIMRICore (r1.3 - 12 Jun 2007 - KarenCrawford) |
ADNI MRI CoreMRI Core News | ||||||||
| Line: 8 to 8 | ||||||||
|---|---|---|---|---|---|---|---|---|
| Deleted: | ||||||||
| < < |
FAQsQ: Why are there cases where the subject has a 6 month scan but no baseline scan? A: The subject's 1.5T screening scan becomes baseline if the subject is approved for the study. | |||||||
Topic permissions
| ||||||||
| <<O>> Difference Topic ADNIMRICore (r1.2 - 22 May 2007 - KarenCrawford) |
ADNI MRI CoreMRI Core News | ||||||||
| Line: 8 to 8 | ||||||||
|---|---|---|---|---|---|---|---|---|
| Added: | ||||||||
| > > |
FAQsQ: Why are there cases where the subject has a 6 month scan but no baseline scan? A: The subject's 1.5T screening scan becomes baseline if the subject is approved for the study. | |||||||
Topic permissions
| ||||||||
| <<O>> Difference Topic ADNIMRICore (r1.1 - 07 May 2007 - KarenCrawford) |
| Line: 1 to 1 | ||||||||
|---|---|---|---|---|---|---|---|---|
| Added: | ||||||||
| > > |
ADNI MRI CoreMRI Core NewsThe ADNI MRI core has uncovered a potential concern about the accuracy of image scaling in the frequency encode direction on Philips 1.5T systems. This problem is currently under investigation. For the time being, we suggest that investigators use this particular data judiciously. This message will be updated as more information becomes available.Topic permissions
| |||||||
|
Revision r1.1 - 07 May 2007 - 18:51 - KarenCrawford Revision r1.8 - 05 Jun 2008 - 18:38 - PaulaBritson |